Serial Port Asynchronous Operations

Contents

Introduction

This demo explores asynchronous read and write operations using a serial port object.

The information obtained for this demonstration was prerecorded. Therefore, you do not need an actual instrument to learn about asynchronous serial port read and write operations. The instrument used was a Tektronix TDS 210 oscilloscope.

Functions and Properties

These functions are associated with reading and writing text asynchronously:

FPRINTF        - Write text to instrument.
READASYNC      - Asynchronously read bytes from an instrument.
STOPASYNC      - Stop an asynchronous read and write operation.

These properties are associated with ASCII read and write asynchronous operations:

BytesAvailable - Indicate the number of bytes available in
                 the input buffer.
TransferStatus - Indicate what type of asynchronous operation
                 is in progress.
ReadAsyncMode  - Specify whether data is read continuously
                 in the background or whether you must call
                 the READASYNC function to read data
                 asynchronously.

Additionally, you can use all callback properties during asynchronous read and write operations.

Synchronous Versus Asynchronous Operations

The object can operate in synchronous mode or in asynchronous mode. When the object is operating synchronously, the read and write routines will block the MATLAB command line until the operation has completed or a timeout occurs. When the object is operating asynchronously, the read and write routines will return control immediately to the MATLAB command line.

Additionally, you can use callback properties and callback functions to perform tasks as data is being written or read. For example, you can create a callback function that notifies you when the read or write operation has finished.

Creating a Serial Port Object

To begin, create a serial port object associated with the COM2 port. The oscilloscope is configured to a baud rate of 9600, 1 stop bit, a carriage return terminator, no parity, and no flow control.

>> s = serial('COM2');
>> set(s, 'BaudRate', 9600, 'StopBits', 1)
>> set(s, 'Terminator', 'CR', 'Parity', 'none')
>> set(s, 'FlowControl', 'none');
>> s
  Serial Port Object : Serial-COM2
  Communication Settings
     Port:               COM2
     BaudRate:           9600
     Terminator:         'CR'
  Communication State
     Status:             closed
     RecordStatus:       off
  Read/Write State
     TransferStatus:     idle
     BytesAvailable:     0
     ValuesReceived:     0
     ValuesSent:         0

Connecting the Serial Port Object to Your Instrument

Before you can perform a read or write operation, you must connect the serial port object to the instrument with the FOPEN function. If the serial port object was successfully connected, its Status property is automatically configured to open.

>> fopen(s)
>> get(s, 'Status')
ans =
open

Reading Data Asynchronously

You can read data asynchronously with the serial port object in one of these two ways:

The FSCANF, FREAD, FGETL, and FGETS functions are used to bring the data from the input buffer into MATLAB. These functions operate synchronously.

Reading Data Asynchronously -- Continuous ReadAsyncMode

To begin, read data continuously.

>> set(s, 'ReadAsyncMode', 'continuous')

Now, query the instrument for the peak-to-peak value of the signal on channel 1.

>> fprintf(s, 'Measurement:Meas1:Source CH1')
>> fprintf(s, 'Measurement:Meas1:Type Pk2Pk')
>> fprintf(s, 'Measurement:Meas1:Value?')

Since the ReadAsyncMode property is set to 'continuous', the object is continuously asking the instrument if any data is available. Once the last FPRINTF function completes, the instrument begins sending data; the data is read from the instrument and is stored in the input buffer.

>> get(s, 'BytesAvailable')
ans =
   15

You can bring the data from the object's input buffer into the MATLAB workspace with FSCANF.

>> data = fscanf(s)
data =
4.1199998856E0

Reading Data Asynchronously -- Manual ReadAsyncMode

Next, read the data manually.

>> set(s, 'ReadAsyncMode', 'manual')

Now, query the instrument for the frequency of the signal on channel 1.

>> fprintf(s, 'Measurement:Meas2:Source CH1')
>> fprintf(s, 'Measurement:Meas2:Type Freq')
>> fprintf(s, 'Measurement:Meas2:Value?')

Once the last FPRINTF function completes, the instrument begins sending data. However, since ReadAsyncMode is set to manual, the object is not reading the data being sent from the instrument. Therefore, no data is being read and placed in the input buffer.

>> get(s, 'BytesAvailable')
ans =
    0

The READASYNC function can asynchronously read the data from the instrument. The READASYNC function returns control to the MATLAB command line immediately.

The READASYNC function takes two input arguments. The first argument is the instrument object and the second argument is the SIZE, the amount of data to be read from the instrument.

The READASYNC function without a specified SIZE assumes SIZE is given by the difference between the InputBufferSize property value and the BytesAvailable property value. The asynchronous read terminates when

An error event will be generated if READASYNC terminates due to a timeout.

Since all the data has been sent when the READASYNC function is called, no data will be stored in the input buffer and the data is lost.

>> readasync(s);
>> s.BytesAvailable
ans =
    0

It is important to remember that when the serial port object is in manual mode (the ReadAsyncMode property is configured to manual), data that is sent from the instrument to the computer is not automatically stored in the input buffer of the connected serial port object. Data is not stored until READASYNC or one of the blocking read functions is called.

Manual mode should be used when a stream of data is being sent from your instrument and you only want to capture portions of the data.

Defining an Asynchronous Read Callback

Now, configure our serial object to notify us when a terminator has been read.

>> set(s, 'ReadAsyncMode', 'continuous')
>> set(s, 'BytesAvailableFcn', {'dispcallback'})

Note, the default value for the BytesAvailableFcnMode property indicates that the callback function defined by the BytesAvailableFcn property will be executed when the terminator has been read.

>> get(s, 'BytesAvailableFcnMode')
ans =
terminator

The Callback Function

The M-file callback function DISPCALLBACK is defined below.

function dispcallback(obj, event)
%DISPCALLBACK Display event information for the specified event.
%
%    DISPCALLBACK(OBJ, EVENT) a callback function that displays
%    a message which contains the type of the event, the name
%    of the object which caused the event to occur and the
%    time the event occurred.
%
%    See also INSTRCALLBACK.
callbackTime = datestr(datenum(event.Data.AbsTime));
fprintf(['A ' event.Type ' event occurred for ' obj.Name ' at ' callbackTime '.\n']);

Using Callbacks During an Asynchronous Read

Now, query the instrument for the period of the signal on channel 1. Once the terminator is read from the instrument and placed in the input buffer, DISPCALLBACK is executed and a message is posted to the MATLAB command window indicating that a BytesAvailable event occurred.

>> fprintf(s, 'Measurement:Meas3:Source CH1')
>> fprintf(s, 'Measurement:Meas3:Type Period')
>> fprintf(s, 'Measurement:Meas3:Value?')
A BytesAvailable event occurred for Serial-COM2 at 29-Dec-1999 17:16:36.
>> get(s, 'BytesAvailable')
ans =
   13
>> data = fscanf(s, '%c', 13)
data =
1.0019999E-3

Note: the last value read is the carriage return (13):

>> real(data)
ans =
Columns 1 through 12
  49    46    48    48    49    57    57    57    57    69    45    51
Column 13
  13

Stopping an Asynchronous Read Operation

Now suppose that halfway through the asynchronous read operation, you realize that the signal displayed on the oscilloscope was incorrect. Rather than waiting for the asynchronous operation to complete, you can use the STOPASYNC function to stop the asynchronous read. Note that if an asynchronous write was in progress, the asynchronous write operation would also be stopped.

>> set(s, 'BytesAvailableFcn', '');
>> fprintf(s, 'Curve?');
>> get(s, {'TransferStatus', 'BytesAvailable'})
ans =
  'read'    [146]
>> stopasync(s);
>> get(s, {'TransferStatus', 'BytesAvailable'})
ans =
  'idle'    [186]

The data that has been read from the instrument remains in the input buffer. You can use one of the synchronous read functions to bring this data into the MATLAB workspace. However, since this data represents the wrong signal, the FLUSHINPUT function is called to remove all data from the input buffer.

>> flushinput(s);
>> get(s, {'TransferStatus', 'BytesAvailable'});
ans =
  'idle'    [0]

Writing Data Asynchronously

You can perform an asynchronous write with the FPRINTF or FWRITE functions by passing 'async' as the last input argument.

Defining an Asynchronous Write Callback

In asynchronous mode, you can use callback properties and callback functions to perform tasks as data is written. For example, configure the object to notify us when an asynchronous write operation completes.

>> set(s, 'OutputEmptyFcn', {'dispcallback'});
>> fprintf(s, 'Measurement:Meas3:Value?', 'async')
A OutputEmpty event occurred for Serial-COM2 at 29-Dec-1999 17:25:48.

Stopping an Asynchronous Write Operation

If the STOPASYNC function is called during an asynchronous write operation, the asynchronous write will be stopped and the data in the output buffer will automatically be flushed. If an asynchronous read operation was also in progress, it would be stopped as well. However, the data in the input buffer would not be flushed automatically.

Cleanup

If you are finished with the serial port object, disconnect it from the instrument, remove it from memory, and remove it from the workspace.

>> fclose(s)
>> delete(s)
>> clear s